This product is mainly used for testing the thermophysical properties of high-temperature liquid samples, including density, conductivity, liquidus temperature, and surface tension. The unique furnace design in this product not only serves as an inert atmosphere protection, but also facilitates the conversion and measurement of different parameter tests quickly and conveniently.
test temperature | RT-1000℃(The maximum temperature of the furnace body is 1350 ℃) | ||||
Density | Archimedes | precision | 1% | reproducibility | 1% |
Conductivity | Continuous transformation of conductivity cell constant method | precision | 5% | reproducibility | 5% |
liquidus temperature | Step cooling curve method | precision | ±1℃ | ||
surface tension | Maximum bubble method | precision | 3% | reproducibility | 3% |
The test sample is stable, non volatile, with low viscosity, uniform in liquid state, and no precipitation, such as various nitrates, chlorides, fluorides, carbonates, and some liquid metals.